Precision in an anesthesia ventilator, from conventional ventilation to advanced modes, including 3 modes: IPPV; A/C; SIMV.
Description

S6100D Aeonmed Drager Mindray Anesthesia Machine

S6100D Aeonmed Drager Mindray Anesthesia Machine

S6100D Aeonmed Drager Mindray Anesthesia Machine

S6100D Aeonmed Drager Mindray Anesthesia Machine

 

 

Features

7” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.

High precision flow meter, instantly know the fresh gas flow to your patient.

Integration breathing circuit design, ensure easy operating and keep tidy.

Multiple working modes such as volume control and pressure limit, adapt to wide range patient.

Vaporizer with temperature, flow compensation and self-lock function (S6100D can only equipped with one vaporizer), keep safety anytime.

Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;

Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.

Vital sign monitor and Anesthetic gas monitor are optional.

 

Safety

Three level alarming system, visual and sound alarm information.

With lots of alarming, reminding and protection functions.

Advanced power management control technology.

Low O2 pressure alarm and N2O cut-off protection.

 

 

Specifications

          

Ventilation modes:

IPPV; A/C; SIMV;

Ventilator parameter ranges

 

Flow meter

O2(0.1-10L/min)

 

N2O(0.1-10L/min)

 

AIR(0.1-10L/min)

Rapid oxygen supply

35L/min-75L/min

Tidal volume(Vt)

0, 20 mL ~ 1500 mL

Frequence(Freq)

1bpm ~ 100 bpm

I/E

4:1~1:8

PEEP

0cmH2O ~ 30 cmH2O

Pressure triggering sensitivity(PTr)

-20 cmH2O ~ 0 cmH2O(Based on PEEP)

Flow trigger sensitivity (FTr)

0.5 L/min ~ 30 L/min

Pressure control (PC)

5 cmH2O ~ 60 cmH2O

SIGH

0(off)  1/100 ~ 5/100

Apnea Ventilation

OFF,5 s ~ 60 s

Pressure Limit

20 cmH2O ~ 100 cmH2O

Monitored Parameters

 

Frequence(Freq)

0 /min ~ 100 /min

Tidal volume(Vt)

0 mL ~ 2000 mL

MV

0 L/min ~ 100 L/min

Oxygen concentration

15 % ~ 100 %

 

Graphical display:

Oxygen concentration

 

Size

Wooden case packing size:L 560*W 820*H 1450mm , GW:104KG ; NW:64KG

 

Alarm and protection 

No tidal volume

≤5Ml within 6s

High oxygen concentration alarm

Low oxygen concentration alarm

19%-100%

18%-99%

High Airway  pressure alarm

Low Airway  pressure alarm

High Minute Volume alarm

Low Minute Volume alarm

Continuous Pressure alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

Adult(5L/min-20L/min) Paed(1L/min-15L/min

0-10L/min)

(PEEP+1.5kPa) over 16s

Suffocation warning

5s-60s no spontaneous ventilation

The maximum limited pressure 

 

Fan error

Oxygen deficit

 

Show on screen

Show on screen

Working conditions

 

Gas source

O2,N2O

Pressure

280kPa-600kPa

Voltage

100-240V

Power frequency

50/60Hz

Input power

40VA

Vaporizer